The instrument basically consists of a miniature furnace with Auto controlled heating rate arrangement, an air circulating arrangement and Digital temp. Indication with necessary controls. When the Start key is pressed, heating starts. There is Auto detection of Opto-coupled light transmission system , When sample melts , photodiode detects the signal and Melting Temperature put on hold / latched on display automatically. The furnace becomes off and the cooling fan ON simultaneously. The instrument gives audio visual indications. The Melting Temp. of the sample remain displayed on the panel till the RESET switch is operated.
Not Suitable for Dark Colored Samples. Opto-coupled light transmission based, Sample required to be transparent at the time of melting.
OPERATION : Fully Automatic Sample Fill and Forget : All the selection of programming is fully protected from power failure including HOLD action of melting temperature. Once the start action is initiated, it will reach to find out the melting point of sample, even multiple time power failure.
NU-138E Melting Point Apparatus Automatic:
Temp. Range Ambient to 350C
Display 16X2 lines alphanumeric display
Resolution 0. 1C
Temp. Sensor Grade A PT-100, Platinum Sensor
Sample Capacity One Sample , Capillary Dimension 1.4 mm Temp. Control Microprocessor controlled closed loop PID
Ramp Rates 2C to 10C/minute
(Selectable Auto-Low-High),
Operation Auto determination of Melting Point
Serial Interface RS-232 C to PC
Memory 100 Samples with Location
Power Supply 230 V 10% AC, 50 Hz.